Malfunction &
Maintenance
Malfunction & Maintenance
//
TriBox3
6.3 Troubleshooting
If the TriBox3 cannot be operated as described in the manual or if it displays other abnormalities, please first
make sure that it is not damaged.
If the possibility of damage can be excluded, it is possible that the operating system is not working properly. In
this case, reboot the system. This is particularly true when a new sensor is not immediately detected.
The “Info” button opens a window with the contact data of the nearest contact in case of problems
or general inquiries. Before contacting the contact person, please read chapter 6 in full to ensure
you have the right information for efficient troubleshooting.
6.3.1 Changing the Fuse
The Tribox3 uses the following fuses:
Fuse F1
1A, 250V, time-lag, 5x20mm; item no. 00P100009
Fuse F2 and F3
4A, 125V, fast-blow, SMD without holder; item no. 00P100008
Fuse F4 to F14
1A, 125V, fast-blow, SMD without holder; item no. 00P100007
Fuse F1
Fuses F2 and F3
If you need to replace a defective fuse, first unplug the power supply and
use only fuses of the same type. Make sure you have completely eliminated
the error before putting the device back into operation.
WARNING
Danger to life due to electric shock! Due to the free mains voltage in the unit,
installation work may only be carried out by qualified personnel who are
authorised to do so on the basis of their training. The relevant safety and VDE
regulations must be observed. Before opening the unit, disconnect it from the
power supply and secure it against being switched on again.
DANGER