39
D01-051en202111 Manual TriBox3
Advanced
Use
Malfunction & Maintenance
Commis- sioning
Introduction
General
Information
FA
Q
Technical Data
W
arranty
Customer
Service
Contact
Keyword
Index
Accessories
Use
TriBox3
//
Use
Recommendations for timer and cleaning intervals*
Measurement
interval***
Cleaning interval
Cleaning time
Pause before mea-
surement
minimum**
30 s
30 s
5 s
5 s
typical**
2 min
15 min
10 s
5 s
maximum
1 day
6 hours
20 s
5 min
* with 10-meter TriOS compressed air hose 4/6 mm
** depending on the type of sensor
*** see chapter 3.4.4 “Options”.
4.6 Data Storage