26
CODE
NAME
t 1
First Test Weight Value
C 1
First Span Calibration Value
t 2
Second Test Weight Value
C 2
Second Span Calibration Value
t 3
Third Test Weight Value
C 3
Third Span Calibration Value
Calibration Value Entry Table
mV/V Span Calibration (F52)
This is an alternative calibration method and is less accurate yet does not require use of known
weights. This method does not affect, change or erase any of the digital scale system calibration
values, so it can be used in a pinch as needed to get your scale system up and running.
1. Press and hold the MENU/OFF key to switch off the digital indicator.
2. Press and hold the ON key until the digital indicator beeps and starts to boot up.
3.
During the countdown phase (“555555, 444444”, etc.) press and hold the MENU/OFF
key until the “
Set?
” message is displayed.
4.
Press the NET/GROSS button while “Set?” is being displayed. The digital indicator dis-
plays
“– F –
“.
5. Scroll down once using the ZERO key to enter
the “Setup” menu. The screen shows
“
F 1
”.
6. Press the TARE (left) key once to reach the "
F 52
" screen, and then scroll down one more
level using the ZERO key to arrive at the "
F 52-1
" screen.
7. Populate the F52 sub-parameters per the Table following. Be sure to press the
NET/GROSS key to save each new value.
8. When finished press the UNITS (up) key two times to reach the
“– F –
“screen and then
press MENU/OFF key once to exit and return to normal operating mode.