3-45
●
HOAs [4mm]:
Displays the total higher-order aberrations within a
Φ4-mm range.
●
SA [6mm]:
Displays any spherical aberration within a Φ6-mm
range.
Refer to “3.6.4 Quantitative index” for details of
quantitative index.
The following quantitative index will be the reference for
IOL formulae and selection of IOL type.
AvgK/Real AvgK : Reference value when applying an IOL formula. Reference value when
determining whether to use a special IOL formula such as that for LASIK
Real CYL
: Reference value when applying Toric IOL
SA
: Reference value when applying a spherical/aspherical lens.When this
value is too small, significant effects may not be seen even if applying an
aspherical lens.
HOAs
: Reference value when applying a premium IOL. When this value is too
large, significant effects may not be seen even if applying a premium IOL.
The quantitative index listed above is color-coded
according to applicable thresholds. The threshold can
be set on the Setting screen.
(Refer to “3.9 System Setup [IOL Screenig tab].”)