3-10
3.3.3 Selecting the capture mode
The ring topo mode (for capturing a corneal shape) and
slit mode (for capturing a cross-section of the anterior
chamber) are available.
When changing the capture mode of the instrument,
pay attention to prevent any moving parts from hitting
the patient’s eye or nose. Also, be sure that the patient
does not place their hands or fingers near any moving
section of the instrument. Doing so may cause injury.
1) Use the joystick to move the head toward the
physician, to the point farthest from the patient.
2) Press the MODE button (1) on the membrane button
panel (Fig. 1) to change modes.
Touching the “Mode” button (2) on the measurement
screen (Fig. 2) of the main unit also changes modes.
(Fig. 1)
(1)
(2)
(Fig. 2)
Physicians's side