3-18
b) 2D scan
The preset number of images of a cross-section of the
anterior chamber are captured in series from the specified
angle.
1) Press the “Temp” button (1) to open the slit mode
temporary screen (Fig. 2).
2) Select “2D scan” (2) and specify the number of images
to be captured (3).
3) Touching the “Exit” button (4) displays the 2D scan
screen (Fig. 3) for 2D capturing.
4) Specify the angle for capturing images. Touch a
number displayed on the screen (5) to turn the angle
indicator bar (6) to the selected angle.
Use the capturing angle adjustment buttons (7) to
finely adjust the angle.
5) After determining the capturing angle, perform
alignment and then capture images in the same way
as 3D scan. (Refer to “3.4.4 a) 3D scan.”)
(1)
(Fig. 1)
(2)
(4)
(3)
(Fig. 2)
(6)
(5)
(7)
(Fig. 3)