i-2
Do not place water or chemicals on the instrument. Any water or chemicals
entering the instrument may cause an electric shock or failure.
This instrument is a diagnostic/capturing device specifically designed for
ophthalmology. Never use the instrument for other purposes.
When operating this instrument connected to other devices not described in
this document, only use devices that satisfy IEC60601-1 or equivalent safety
requirements or that conform to IEC60950 and whose power source is insulated
with the attached Isolation transformer, in order to satisfy the safety requirements
regarding the medical equipment.
The total rated input of devices to be connected to the Isolation transformer is
less than 500 VA. Use the transformer to supply power only to devices intended
to comprise part of the system.
Do not connect the power source of the personal computer, which is connected
to this instrument for operation, directly to the socket on the wall. Otherwise, an
electric shock or failure may occur.