6-2
Blown fuse
Check that fuses are not blown. If blown,
replace the fuse (refer to “5.5.1 Fuses”).
If the new fuse is blown again, the
instrument may be faulty. Contact our
local distributor to request inspection
and/or repair.
● The ring cone lamp or fixation lamp does not light in
ring topo mode.
The Capture screen is not open on the
personal computer.
Check that other screens (Map screen,
etc) are not shown on the personal
computer.
Even though the Capture screen is open,
the joystick button needs to be pressed to
turn on the lamp.
Press the joystick button to turn on the
lamp.
Similarly, press the joystick button if
Time Out has occurred.
● OD/OS displayed on the main unit is not changed
even though the head is moved to change the eye for
measurement.
The sensor that detects the right and left
positions of the head is not activated.
Stop using the instrument and contact
our local distributor.
?
?
?
?