6.
CARE OF THE INSTRUMENT
77
1.
Open the retort lid. If the retort is hot, wait until it cools
down.
2.
Moisten the tip of the level sensor cleaning brush with
alcohol.
Figure 6-D
3.
Hold the brush with the tip downward. Insert the tip into
the underside of the sensor probe as shown in the photo
below.
NOTE:
The level sensors are located on the front wall
of the retort.
Figure 6-E
Insert the brush into the sensor hole
Figure 6-F
4.
Turn the brush 90 degrees to the right or left to place
the brush tip into the groove of the sensor probe as
shown in the photo below. If the brush is caught when
you pull it out gently, the brush tip is set in the groove
of the sensor probe.
Groove of the sensor probe
Figure 6-G
The brush is caught when pulled out
Figure 6-H