66
4.4.4
Removing AutoTEC Magazines
When removing the magazines from the retort, to prevent
misalignment or dislodging of cassettes, gently grasp the
magazines from the top and rotate to an upright position with
correct side up, as show below (Figure 4-M).
Figure 4-M
4.5
Cleaning
After a processing run the Clean Cycle Start Screen displays
(Figure 4-N). The Clean Cycle Start Screen consists of three
tabs, Clean Retort, Warm Water Flush and Rinse Cycle.
When a program is selected the lamp icon illuminates.
Figure 4-N
4.5.1
The Clean Retort Tab
On the Clean Retort Tab information pertaining to the clean
cycle is displayed.
•
Touch the
Edit Clean Programs
button to edit the
current cleaning program.
•
Touch the
Modify
button to change the
Dechloroform cycle to In Use or Not In Use.
•
Touch the
Modify
button to change the warm-up
time. The setting is 0-15 minutes.
•
Touch the
Solution Manager
button to enable or
disable the solution manager function. If the
solution manager is enabled the lamp icon is
illuminated and the solution manager icon is
displayed in the upper right of the screen.
•
Touch the
Drying Cycle
button to enable or
disable the drying function. If the drying cycle is
enabled the lamp icon is illuminated and the drying
cycle icon is displayed in the upper right of the
screen.