Setting up Exchange Solutions:
Reagents
This option allows the user to access the solution exchange
operations.
1.
From the Process Menu, touch the Exchange Solution
button.
•
The Exchange Solution screen displays (Figure
3-VVV).
Figure 3-VVV
2.
Touch to select one of six tabs: Group Exchange, Single
Solution, Automatic Transfer, Paraffin Drain, Reservoir
Fill, and Defatting Solution.
To perform a group exchange:
3.
Under the Group Exchange Tab, touch a solution group
to be exchanged.
4.
Select the desired type of exchange: External Drain,
External Fill, Fill from Bulk Reservoir, and Drain & Fill
from Bulk Reservoir.
NOTE:
Fill from Bulk Reservoir and Drain & Fill
from Bulk Reservoir are not available for the
solution groups where the bulk reservoir supply has
not been selected.
5.
Touch the
View as a List
button to view the stations
as a list.
•
The Solution List screen displays (Figure 3-WWW).
Figure 3-WWW
6.
Touch
Exit
to return the process menu.
7.
Touch
Start
to begin the process.
•
The Exchange Start Confirmation Window displays
(Figure 3-XXX).
Figure 3-XXX
8.
Touch
Yes
to start the Solution Group Exchange.
9.
Touch
No
to return back to the Exchange Solution
Screen.
•
The Exchange Solution Overview Monitor Screen
displays (Figure 3-YYY).