44
•
The Input Hour Window is displayed (Figure 3-KKK):
Figure 3-KKK
4.
Use the touch keypad to enter the hour. The settings are
1-12 in the 12-hour format and 0-23 in the 24-hour
format.
•
Touch
Enter
to save the settings.
•
Touch
Clear
to erase the settings.
•
Touch
Cancel
to exit without saving.
5.
Touch the
Modify
button under the minutes field to
modify the time at which processing will be completed.
•
The Input Date Minutes is displayed (Figure 3-LLL).
Figure 3-LLL
6.
Use the touch keypad to enter the minutes. The setting
is 0-59.
•
Touch
Enter
to save the settings.
•
Touch
Clear
to erase the settings.
•
Touch
Cancel
to exit without saving.
7.
Touch the AM/PM button to select either “am” or “pm”.
When the 24-hour format is used, this button is not
displayed.
•
Touch
Save
to save the predicted end time
settings.
•
Touch
Cancel
to exit without saving.
3.4.6
Copying Existing Programs
Existing programs can be used as a template to create a
new program. When a program is copied, program steps and
settings are duplicated to create a new program.
1.
From the Edit Menu, touch the
Edit Program
button.
•
The Edit Program screen displays.
2.
From the Program List, select to highlight the program
to be copied and then touch the
Copy
button.
•
The Copy Program screen displays.
3.
Select to highlight the program to copy the information
to or touch
Add Program
to add a new program to the
list.
Solution Manager
The system performs “automatic solution replenishment” to
supply a new solution to a station where the system
detected a shortage of solution during tissue processing or
when retort cleaning is in progress.
This function is valid when the solution level safeguard
setting is “In Use” in the system setup. The automatic
solution supply operation may be performed during tissue
processing, retort cleaning or rinsing. Two types of solution
management are available:
Solution Manager 1
Low Fluid
– When the system detects a shortage during the
pump-in phase during processing or cleaning, solution is
supplied from the next station in the same group until the
solution safeguard level sensor (2.7L for 150 cassettes/3.5L
for 300 cassettes) is activated. If the next station cannot be
used, new solution is supplied from the linked bulk reservoir.
NOTE:
For the retort cleaning, this function
supplies a solution from the bulk solution reservoir
linked to the retort until the upper limit level sensor
(4.2L) is turned on. If the bulk solution reservoir is