16
6.
Touch the
Process Menu
tab.
7.
Touch
Exchange Solution
button.
8.
Touch the
Single Solution
tab.
9.
Touch the solution to be filled.
10.
Touch the
External Fill
button.
11.
Touch the
View as a List
button to view the solution
list and then touch
Exit
to continue.
12.
Choose the volume of the solution for the station chosen
(150 cassettes, 300 cassettes, or cleaning).
13.
Touch
Start
to begin the solution exchange or
Exit
to
exit the screen.
•
If desired, the drain/fill port can be rinsed using the
Rinse function.
14.
Exchange Confirmation:
•
Touch
Yes
if connection is secure.
•
Touch
No
if connection is not secure and then
connect the drain/fill hose.
15.
Start Confirmation:
•
Touch
Yes
when ready to exchange.
16.
Start Notice:
•
Touch
Exit
.
17.
End Notice:
•
Touch
Exit
.
Installing the Waste Paraffin Reservoir
Liner
1.
Insert black Anti-Adhesion sheet into Paraffin Waste
reservoir (Figure 2-H).
2.
Take one liner out from the package.
3.
Spread out the folded liner (Figure 2-H).
Figure 2-G
Figure 2-H
4.
Unfold the stiffened portion of the liner (Figure 2-I).
Figure 2-I