The Exchange Hold (Solution Group Exchange) screen
is displayed (Figure 3-AAAA).
Figure 3-AAAA
3.
Touch
Resume
to resume the Solution group exchange.
4.
Touch
Abort
to abort the process (Figure 3-BBBB).
Figure 3-BBBB
•
Touch
Yes
to abort.
•
Touch
No
to disregard request.
To perform a single solution exchange:
1.
Under the Single Solution Tab, touch the
solution
to be
exchanged.
2.
Touch the
Solution Level
button below the selected
solution to choose the correct solution level (150
cassettes (2.7 L), 300 cassettes (3.5 L), Cleaning Level
(4.2 L)).
3.
Select the desired type of exchange: External Drain,
External Fill, Fill from Bulk Reservoir, and Drain & Fill
from Bulk Reservoir.
NOTE:
Fill from Bulk Reservoir and Drain & Fill
from Bulk Reservoir are not available for the
solution groups where the bulk reservoir supply has
not been selected.
4.
Touch Exit to return to the process menu.
To perform an independent automatic transfer:
Independent automatic transfer can rotate up to four
different groups of solution, including paraffin, during one
solution exchange operation. When using this function, it
does not matter whether or not the stations of a particular
solution group to be rotated are linked to the bulk solution
reservoir. For a solution group which is not linked to the bulk
solution reservoir, the automatic transfer is completed after
emptying the solution bottle at the last station of that
solution group. In this case, the last station needs to be filled
with a new solution manually or by using the
Fill
function of
the solution exchange feature.
The independent automatic transfer allows the operator to
designate the solution volume to be rotated. When a solution
in the source station is more than the solution volume
designated for the destination station, only the designated
solution volume is rotated to the destination station. When
the source station is not full enough to supply the designated
volume to the destination station, a solution is rotated first
from the source station and then from the next station in
order to fill the destination station. When the source station
is the last station of the solution group, an appropriate
amount of solution is supplied from the bulk reservoir.
1.
Touch the solution group button to select the group to
automatically rotate.
NOTE:
The first station of the group must be
emptied prior to the start of operation.
NOTE:
If paraffin is being rotated, the paraffin
waste reservoir must be emptied prior to operation.
2.
Touch the
Level
button below the station group to
choose the correct solution level (150 cassette level (2.7
L), 300 cassette level (3.5 L), or cleaning level (4.2 L).
3.
Touch
View
as List to view as a list.
4.
Touch
Start
to begin the process.
5.
Touch
Exit
to return to the process menu.