© 2011 Thorlabs
139
Application Note
X= SUM [x * p (x,y)] / I
Y= SUM [y * p (x,y)] / I
where:
p(x,y) intensity at location (x,y);
I
total intensity;
SUM of pixels taken taken over entire area
AD Saturation
Saturation level of the instrument's AD converter.
Total Power
.
Effective Area
Area of an ideal flat top beam with same peak intensity:
PixelArea
ity
PeakIntens
ls
SumAllPixe
A
ef f
*
For calculation into mm² , PixelArea is equal to 6,45*6,45 µm²
Peak Density
Power on peak pixel divided by its area
59