Operating the Beam Profiler
The panels of the
Calculation To Actual Z Position
and
Results
are described in
section.
3.5.7.2
Divergence Measurement Settings
For a successful and reliable measurement it is essential to adapt the settings of the
software.
Click on
to enter the
Divergence Measurement Settings
.
Beam Width
The Beam Width can be selected from the two already known methods.
The calculation of the
4 Diameter
follows the ISO 11146-1:2005 standard.
The
Approximated Ellipse
beam width is based on the 13.5% clip level. At the
beginning of a measurement a reference angle is determined by averaging over 10
frames. This angle is then used to evaluate all following frames and ellipses.
Measurement Parameters
Timeout
is the max. waiting time for a valid camera image. This is to allow the
Autoexposure function to obtain a valid camera image
Wavelength
is not relevant for divergence measurement.