Installing the iN10/iN10MX
This document contains confidential or proprietary information of Thermo Fisher Scientific. Neither this
document nor the information therein is to be reproduced, distributed, used or disclosed, either in whole or
in part, except as specifically authorized by Thermo Fisher Scientific.
8/3/2020
Page 20 of 61
32.5.
Open the Stage Control screen.
From the
Window
drop down
menu choose
Stage Control
.
32.6.
Initialize the X-Y Stage.
Click
Initialize XY
to begin initialization.
32.7.
Check the centering of the stage.
At the right front corner of the
stage, check that all of the stage plates are flush. If so, continue
with step 32.11, if not continue with 32.8.
32.8.
Align Stage plates.
From the
Window
drop down menu choose
Joystick Control
. Using the X-Y stage controls, move the stage
until all of the stage plates are flush with each other. Close the
Joystick Control
screen when complete.