33.5.
Configure ATR.
From the
Window
drop down menu choose
User
Configuration
. Check
ATR is installed
.
If you do not have an ATR, continue with step 34.
34.
Verify aperture centering parameters.
From the
DDiag\Protected Content
menu,
open
Configure iN10.
Verify the Aperture Centering parameters are a
four-digit number (three digits for earlier systems as depicted below) and not
zero or appear to be corrupted. If OK, continue with step 35 otherwise
proceed with the following steps.