Installing the iN10/iN10MX
This document contains confidential or proprietary information of Thermo Fisher Scientific. Neither this
document nor the information therein is to be reproduced, distributed, used or disclosed, either in whole or
in part, except as specifically authorized by Thermo Fisher Scientific.
8/3/2020
Page 53 of 61
46.12.
Remove the ATR crystal from the IR beampath.
Pull the ATR slide
out until it clicks in the first detent and is out of the IR beampath.
46.13
Insert one of the polycarbonate slides from the customer tool kit
onto the ATR sample holder.
46.14
Change system parameters.
From the
Experiment Setup
window set
the
Collection
parameters as follows:
Collection Mode = ATR
46.15
Move the stage to the polycarbonate window.
Using either the
virtual or manual joystick, move the stage to position the polycarbonate
window under the objective.
46.16
Focus and locate sample area.
Using either the virtual or manual
joystick, focus the Z stage. Find an area on the polycarbonate window
that is clean.