Xcalibur Diagnostics for the PDA Detector
Calibrating the PDA Detector
40
Surveyor PDA Plus Detector Hardware Manual
Thermo Scientific
Figure 37.
Calibration page – Dark Current Calibration area
4. To start the Dark Current Calibration wizard, click
Calibrate
in the
Dark Current
Calibration
area.
The Preconditions page appears (see
Figure 38.
Preconditions page of the Dark Current Calibration wizard
5. Read the Preconditions page and verify that they have been met:
• If the preconditions have been met, click
Next
to proceed with the calibration.
Page two of the Dark Current Calibration wizard appears (see
).
• If the preconditions have not been met, click
Cancel
to exit the wizard. Then prepare
the Surveyor PDA Plus Detector for calibration and begin this procedure again.