Routine Maintenance
Cleaning the LightPipe Flow Cell
88
Surveyor PDA Plus Detector Hardware Manual
Thermo Scientific
Figure 75.
View of the Surveyor PDA Plus Detector with the front doors open
Figure 76.
View of the retaining block and retaining block knob
4. Unscrew and remove the retaining block knob that holds the LightPipe retaining block in
place. Then remove the retaining block (see
5. Being careful not to touch the optical fibers at the ends of the LightPipe flow cell, pull the
flow cell out of the flow cell compartment.
STANDARD FILTER WHEEL
1. : OPEN
2. : HOLMIUM OXIDE FILTER
1
Captive screw
Flow cell access cover
1
Retaining bolt
Retaining block
Retaining block knob
CAUTION
Do
not
touch the ends of the LightPipe flow cell as you remove it from the
flow cell compartment.