![Thermo Scientific Surveyor PDA Plus Hardware Manual Download Page 130](http://html1.mh-extra.com/html/thermo-scientific/surveyor-pda-plus/surveyor-pda-plus_hardware-manual_1105373130.webp)
A
Firmware
Updating the Firmware
110
Surveyor PDA Plus Detector Hardware Manual
Thermo Scientific
Figure 84.
Surveyor Firmware Upgrade Utility, showing the Surveyor PDA Plus option selected
2. Start the communication between the detector and the utility:
a. Under Device Type, click the
Surveyor PDA Plus
option.
b. Click
Connect
to initiate communication between the utility and the Surveyor PDA
Plus Detector.
The status display at the bottom of the Surveyor Firmware Upgrade Utility window
indicates that a connection has been made (see
CAUTION
Do
not
interrupt the firmware download process. Do
not
turn off the
power to the Surveyor PDA Plus Detector or close the Surveyor Firmware Upgrade
Utility while the detector is connected to the Surveyor Firmware Upgrade Utility. If
you lose power to the detector during the firmware download process, contact your
local Thermo Fisher Scientific representative. The firmware board will need to be
reprogrammed.