Thermo Scientific AquaSensors™ DataStick™ Measurement System User Guide
54
ASCII Command to View Cell Constant
Function
Command
Response
Explanation
Get Cell
GCELL
Cell constant
Limits depend on nominal cell
type – 0.1, 1.0, or 5.0
ASCII Command to Set Cell Constant
Function
Command
Explanation
Set Cell
SCELL X
Where X is the desired cell constant
All Set Commands will respond with “OK” if accepted and “ERROR” if not accepted
4 . 3 . 2 . 6
C O N D U C T I V I T Y T E M P E R A T U R E C O M P E N S A T I O N T Y P E
The commands in this section are available in DataStick firmware D3.32 or later. The DataStick firmware version can be obtained using
the GCVSN command.
ASCII Command to View Conductivity Temperature Compensation Type
Function
Command
Response
Explanation
0
The type of temperature compensation used
is
Linear
(default)
1
The type of temperature compensation used
is
Natural Water
View conductivity
temperature
compensation type
GCTCTYPE
2
The type of temperature compensation used
is
None
ASCII Commands to Set Conductivity Temperature Compensation Type
Function
Command
Explanation
Set conductivity
temperature
compensation type
to
Linear
SCTCTYPE 0
The type of temperature compensation used will be Linear and
it will be based on the reference temperature and slope set with
the SCRTEMP and SCCSLOPE commands, respectively
Set conductivity
temperature
compensation type
to
Natural Water
SCTCTYPE 1
The type of temperature compensation used will be
Natural Water
Set conductivity
temperature
compensation type
to
None
SCTCTYPE 2 The type of temperature compensation used will be None
All Set Commands will respond with “OK” if accepted and “ERROR” if not accepted.