Thermo Scientific AquaSensors™ DataStick™ Measurement System User Guide
90
Press the CALIBRATE button to initiate the calibration or click EXIT to return to the main measure screen.
Once the calibration has been initiated it will take between 30 seconds and 2 minutes depending on sensor and temperature
measurement stability. During this period, the abort sequence can be initiated as shown with the active ABORT button.
.
When the point has been successfully calibrated, measured values will be based on the new calibration. The reading will be
in the range of 8 ppm to 10 ppm depending on the atmospheric pressure setting. Consult dissolved oxygen tables for more
specific information on expected readings after an AIR calibration.
Exiting the calibration menu will return the user to the measure screen.
ASCII Command to Initiate an Air Calibration
Function
Command
Explanation
1-Point Sample
CALSAIR
Initiates an AIR calibration for DO or Ozone sensors
All Calibration Commands will respond with “OK” if accepted and “ERROR” if not accepted
The status of the calibration can be reviewed with the CALSTATUS command. When the last number returned is a 1,
the calibration is in process. When the last number returned is a 2, the calibration has been successful. Any other number
indicates a specific failure mode (see appendix).