Thermo Scientific AquaSensors™ DataStick™ Measurement System User Guide
59
Scroll to point 4 and replace the zero coordinates with 100,000µS/cm at 97% concentration. Then click on the STORE button
to save point 4 to temporary memory.
Data for points 5 through 10 should be zero. Data stored in temporary memory can now be permanently stored using
the ACTIVATE STORED TABLE button.
If any error messages are shown, review the data points to be sure they are monotonic, that conductivity increases from
point 1 and that concentration increases or decreases from point 1.
Once the table has been activated, use the EXIT button and return to the main measure menu.
Table entries for specific chemicals are available from Thermo Fisher Scientific Customer Service on request.
The table entries made in this example are shown in Table 2.