Thermo Scientific AquaSensors™ DataStick™ Measurement System User Guide
71
4 . 4 . 2 . 5
C O N D U C T I V I T Y T E M P E R A T U R E C O M P E N S A T I O N T Y P E
The commands in this section are available in DataStick firmware D3.32 or later. The DataStick firmware version can be obtained using
the GCVSN command.
ASCII Command to View Conductivity Temperature Compensation Type
Function
Command
Response
Explanation
0
The type of temperature compensation used
is
Linear
(default)
1
The type of temperature compensation used
is
Natural Water
View conductivity
temperature
compensation type
GCTCTYPE
2
The type of temperature compensation used
is
None
ASCII Commands to Set Conductivity Temperature Compensation Type
Function
Command
Explanation
Set conductivity
temperature
compensation type
to
Linear
SCTCTYPE 0
The type of temperature compensation used will be Linear and
it will be based on the reference temperature and slope set with
the SCRTEMP and SCCSLOPE commands, respectively
Set conductivity
temperature
compensation type
to
Natural Water
SCTCTYPE 1
The type of temperature compensation used will
be Natural Water
Set conductivity
temperature
compensation type
to
None
SCTCTYPE 2 The type of temperature compensation used will be None
All Set Commands will respond with “OK” if accepted and “ERROR” if not accepted.
4 . 4 . 2 . 6
C O N D U C T I V I T Y R E F E R E N C E T E M P E R A T U R E
Select REFERENCE TEMPERATURE from the Configuration pull-down menu to view or change the conductivity
reference temperature.
The setting can be changed by selecting the edit box and typing the desired number. To accept the change, click OK.
To abort the change, click CANCEL.
The default reference temperature for conductivity is 25°C. The system will accept any number between 0 and 100
with up to one decimal place.
Edit the reference temperature by typing a new value in the edit box and clicking OK.