Thermo Scientific AquaSensors™ DataStick™ Measurement System User Guide
70
ASCII Commands to Set Sensor units
Function
Command
Explanation
Set to µS/cm
SSUNITS 0
Sensor data will be reported in µS/cm units
Set to %
SSUNITS 1
Sensor data will be reported in % units
Set to ppm
SSUNITS 2
Sensor data will be reported in ppm units
Set to M
Ω
·
cm
SSUNITS 3
Sensor data will be reported in M
Ω
·
cm units
All Set Commands will respond with “OK” if accepted and “ERROR” if not accepted
4 . 4 . 2 . 4
T E M P E R A T U R E U N I T S
Select TEMPERATURE UNITS from the Configuration pull-down menu to view or change units of measure.
Units shown in the temperature units window are active. To change units or to review options, scroll the selection list. Click on
the desired units of measure to place the new choice in the sensor units window. To accept the change, click OK. To abort the
change, click CANCEL.
Temperature can be displayed in °C or °F. Default units are °C.
ASCII Commands to View Temperature units
Function
Command
Response
Explanation
View Temperature Units
GTUNITS
0
Data reported in °C
1
Data reported in °F
ASCII Commands to Set Temperature units
Function
Command
Explanation
Set °C
STUNITS 0
Sensor data will be reported in °C
Set °F
STUNITS 1
Sensor data will be reported in °F
All Set Commands will respond with “OK” if accepted and “ERROR” if not accepted