Thermo Scientific AquaSensors™ DataStick™ Measurement System User Guide
55
4 . 3 . 2 . 7
C O N D U C T I V I T Y R E F E R E N C E T E M P E R A T U R E
Select REFERENCE TEMPERATURE from the Configuration pull-down menu to view or change the conductivity
reference temperature.
The setting can be changed by selecting the edit box and typing the desired number. To accept the change, click OK.
To abort the change, click CANCEL.
The default reference temperature for conductivity is 25°C. The system will accept any number between 0 and 100
with up to one decimal place.
ASCII Command to View Conductivity Reference Temperature
Function
Command
Response
Explanation
Get Reference
GCRTEMP
A number between 0 and 100
Data reported in °C
ASCII Command to Set Conductivity Reference Temperature
Function
Command
Explanation
Set Reference
SCRTEMP X
Where X is a number between 0 and 100. Units are in °C
All Set Commands will respond with “OK” if accepted and “ERROR” if not accepted
4 . 3 . 2 . 8
C O N D U C T I V I T Y T E M P E R A T U R E C O M P E N S A T I O N S L O P E
Select COMPENSATION SLOPE from the configuration menu to view or change the compensation slope for conductivity
measurements.
The default compensation slope is 2.00%/°C. The system will accept any number from 0 to 9.9 with up to two decimal places.
The setting can be changed by selecting the edit box and typing the desired number. To accept the change, click OK.
To abort the change, click CANCEL.
The reference can be changed to any temperature between 0°C to 100°C with one-tenth degree resolution. Edit the reference
temperature by typing a new value in the edit box and clicking OK.