Thermo Scientific AquaSensors™ DataStick™ Measurement System User Guide
94
When the point has been successfully calibrated, measured values will be based on the new calibration. The change in reading
from the old measured value to the new will be dependent on the temperature filter value that is entered in the configuration menu.
Exiting the calibration menu will return the user to the measure screen.
ASCII Command to initiate a 1-Point Temperature Calibration
Function
Command
Explanation
Temperature
Calibration
CALST1PS
Initiates a 1-point calibration of temperature
All Calibration Commands will respond with “OK” if accepted and “ERROR” if not accepted
The status of the calibration can be reviewed with the CALSTATUS command. When the last number returned is a 1, the calibration
is in process. When the last number returned is a 2, the calibration has been successful. Any other number indicates a specific failure
mode (see appendix).