3-9
When zero is reached or END is pressed, the system displays the following screen :
The word COMPLETE is flashing.
The percentage of error is related to the scale capacity.
Press
NO
returns
MENU 1
without zero changed..
Press
YES
for to modify the zero with the new value. The next screen is shown :
Press
MENU
for reply zero automatic calibration.
Pressing
RUN
returns to menu
RUN
3.8.1.1 Manual
Zero
The
Manual Zero
procedure allows the operator to directly enter the zero constant if known.
Use
ENTER
key for to confirm the new value.
Default:
40000
Min:
0
Max:
120000
The ADV key is only displayed if Auto Zero Tracking optional function is enable. The AZT
function accurately tracks the zero of each scale by calculating an additional zero constant. The
portion of zero due to AZT is not incorporated in the zero constant, but is shown separately.
AUTO ZERO COMPLETE
Error
±
000.00%
Change zero?
YES NO
ZERO CHANGED
New zero
00000
Old zero
00000
RUN MENU
-MANUAL ZERO-
Gross 000.0
Zero
00000
ENTER EXIT (ADV)
Password:Operator