3-14
3.8.2.1.6 Ending an Auto Span Procedure with Test Weights
When is executed an
Auto Span
with test weights use the following procedure:
1.
Press
RUN,
the following screen is displayed:
2.
Press
RUN
Restart the totalizer and returns to menu
RUN
.
3.8.2.1.7
Ending an Auto Span Procedure with R-Cal
When is executed an
Auto Span
with R-Cal use the following procedure:
1.
Press
RUN
2.
R-Cal
relay is de-energized and the display is stead for three second.
3.
Press
RUN
Restart the totalizer and returns to menu
RUN
.
3.8.2.2 Manual
Span
If the span constant is known, the manual span procedure allows the operator to manually
change span.
Use the following procedures:
1.
Press
MENU
until indication
MENU 1
.
the following screen is displayed :
2.
To select
CAL SPAN
the following screen is displayed :
Remove test weights
Bifore return normal
functions
RUN
Menu
-- MENU 1 --
Press menu for others
ZERO SPAN LINEAR
CAL CAL
XXXX = R-Cal, Test Weights
AUTO SPAN XXXX
Scale empty, then press
START
START MANUAL
Summary of Contents for Ramsey MINI CK100