3-24
3.10.3.4 Load Cell Resistance
The resistance value of the bridge of each load cell has to be entered here. The number of
scroll depends on the number of load cells of the system specified on the
MENU 3
.
X
= A, B, C, D, E, o F
N
= 1, 2, 3, 4, 5, o 6
Default:
350
Ohms
Min
: 10
Ohms
Max:
2000
Ohms
Note:
. For each load cell must be inserted the value on the corresponding screen.
3.10.4 Stable
Weight
These Scrolls define the parameters for the stable weight indication. Motion band defines the
range on which the weight should stay in order to be considered stable.
Default:
1
Min:
0
Max:
3
3.10.5
Stable Weight Delay
The motion delay defines how many times this condition should be true before stable weight
indication turns on.
Default:
1
Min:
0
Max:
60
--SC DATA SCROLL 9A -
-
Resistance Load cell N #
350.000 Ohms
Password:Service
-SC DATA SCROLL 11A --
Motion Band division
1
ENTER
Password:Service
-- SC DATA SCROLL 11B -
Motion delay
1.0 sec
ENTER
Password:Service