3-28
Default:
20
sec
Min:
2
sec
Max:
60
sec
3.12
Main Menu 3
MAIN MENU 3
is used for protecting and un-protecting the system using passwords, and to
perform diagnostic and test functions. The diagnostic functions can only be operated after
removing all password protection, and should only be used by experienced technical personnel.
Most test functions are not password protected.
The
PROTECTION
menu only becomes visible after passwords have been defined (see the
DIAGNOSTICS
Menu).
3.12.1
Changing Protection Level
The
MiniCK100
has three protection levels to which specific password are related.
Table 3-2: Protection Levels
Protection Password
Status
NONE
SERVICE
The system is completely unprotected; all data
can be read or changed.
LIMITED OPERATOR
Operator
functions and data are unprotected. All
setup and calibration data are protected except
zero calibrate.
PROTECTED
The system is totally protected, process data can
be read, no change allowed.
A
SERVICE
password is required to access the
NONE
level. An
OPERATOR
or a
SERVICE
password is required to access the
LIMITED
level.
Use the
NONE
key to access the
NONE
protection level. If the current level is not already
NONE
, the
SERVICE
password is required.
-- MAIN MENU 3 -
Press MENU for more
PROT DIAG TEST
-- CAL DATA SCROLL 8C--
Auto Zero Tracking
duration 20 sec
ENTER SCALE#
-- CAL DATA SCROLL 9--
Drift
CHOICE ENTER
Summary of Contents for Ramsey MINI CK100
Page 1: ...1 Operating And Service Manual Ramsey MINI CK100 Weight Indicator MINI CK100 Part No...
Page 4: ......
Page 14: ......
Page 24: ......
Page 80: ......
Page 86: ......
Page 92: ......
Page 94: ......
Page 95: ...C 1 Appendix C Drawings Wiring Diagram CK 0000100 E001C...
Page 96: ......
Page 97: ......