3-18
3.8.3 Linearization
Manual linearization can be accomplished by applying a known test weight(s) or loading the bin
with pre-weighed material and calculation the scale error.
Press
CHOICE
for selections,
YES
to enable, or
NO
to disable linearization. Once enabled,
no linearization is done until the operator manually enters the linearization factors.
Default:
No
Selezioni
: Yes,
No
The linearization factors can be entered with the following screens.
If you enter 1.000 (default value), the load will not be corrected in that portion of the range. A
number lower than 1.000 will reduce the span, while a number larger than 1.000 will increase
the span.
Factor: Default:
1.000
Min:
0.500
Max:
1.500
Then to insert the weight values and corresponding factors, until 5 different points.
(Pressing ACQ for acquired actual weight value). Then to insert until 5 points.
3.9 Setup and Configuration Menu
The following menu, from
MENU 2
to
MENU 5
are used for
SETUP
,
CONFIGURATION
and
DIAGNOSTIC
of the system.
For
MENU 2
selection press SETUP key then ARROW DOWN until the screen
MAIN
MENU
2
. The submenu available are:
•
Display
•
Scale Data
•
Calibrazione Data
- LINEARIZ. 1 -
Linearization
NO
CHOICE ENTER
Password:Service
- LINEARIZATION #1 -
Weight 0,0 kg
Fact. 1,000000
ENTER ACQ FACTOR
Password:Service
Summary of Contents for Ramsey MINI CK100
Page 1: ...1 Operating And Service Manual Ramsey MINI CK100 Weight Indicator MINI CK100 Part No...
Page 4: ......
Page 14: ......
Page 24: ......
Page 80: ......
Page 86: ......
Page 92: ......
Page 94: ......
Page 95: ...C 1 Appendix C Drawings Wiring Diagram CK 0000100 E001C...
Page 96: ......
Page 97: ......