Maintenance and Calibration
Installing the PDM3700 Cover Panel
5-24
PDM3700 Instruction Manual
Thermo Fisher Scientific
Figure 5–32.
Cover panel gasket checkpoints
4.
Install the screws in locations 1 and then 2 and tighten until snug
(
Figure 5–33
). Do not overtighten. These will be tightened to the
proper torque specification at a later step. Press down on back of cover
panel, aligning holes 3 and 4. Put a screw into either location 3 or 4
and again, tighten until snug. Torque the screws to 6-in. lbs,
respectively on location 1 and then 2.
5.
Install the remaining screws into the cover panel to get them started in
a cross-pattern as shown in
Figure 5–33
. Then, tighten until seated in a
cross-pattern as shown in
Figure 5–33
. Finally, torque the screws to 6-
in. lbs in a cross-pattern as shown in
Figure 5–33
.
Figure 5–33.
Cover panel screws
3
7
5
10
4
6
8
9
11
12
13
14
1 2