Troubleshooting
Instrument Diagnostics
Thermo Fisher Scientific
PDM3700 Instruction Manual
6-3
4.
Click on the Diagnostics button on the main WinPDM screen.
The diagnostics runs two different tests. The first series of tests is done with
the pump and heaters off. If failures are detected be sure to note the type
and the reading.
After a short time the computer will prompt to start the second series of
tests. Again, note the type and readings associated with any failures. This
information will assist the service department in diagnosing any problems.
The following is a list of possible causes for various conditions.
Pump and Heaters OFF Test
WinPDM Instrument Diagnostics
Cause
Solution
AIR TEMP
Defective/Disconnected
Sensor
Check connection.
AMBIENT PRES
Defective Sensor
Contact TFS service
AMBIENT TEMP
Defective/Disconnected
Sensor
Check connection.
CLOCK
Defective Component
Contact TFS service
DIFFERENTIAL PRESSURE
Defective Sensor
Contact TFS service
FREQUENCY
Mass Transducer not
installed.
Filter not installed
Mass Transducer
damaged
Install Mass Transducer
Check for filter
Contact TFS service
NO MASS TRANSDUCER
Mass Transducer not
installed
Install Mass Transducer
RH SENSOR
Defective Sensor
Contact TFS service
RH TEMP
Defective Sensor
Contact TFS service
TE TEMP
Defective/Disconnected
Sensor
Confirm Mass Transducer
installed
FLOW RATE (VOLTS)
Board Defective
Contact TFS service
TILT Z
Defective Component
Contact TFS service
BATTERY VOLTS
Charger Off
Battery Low
Charger Defective
Battery Defective
Turn Charger On
Charge Battery
Change Charger
Change Battery