Basic Operation
Stopping a Sample Run
2-20
PDM3700 Instruction Manual
Thermo Fisher Scientific
The Stop Sampling screen contains the following information:
●
(A) Stop Sampling
. This field contains directions on how to stop a
manual sample run. Press the A button to stop the sample run.
●
(B) more options...
This field contains directions on how to display the
various operating screen. When in the Stop Sampling screen, press the
“B” button to display either the Exit screen, View Status screen, or the
Reset Sample 2 screen, depending on the screen from which you have
accessed the Stop Sampling screen.
Users can stop the unit either during the warm-up time or during the
sample run.
1.
In the Warming screen, press both the “A” and “B” buttons
simultaneously to display the Stop Sampling screen.
2.
In the Stop Sampling screen, press the “A” button (“Stop Sampling”
option) to stop the manual sample run and display the Idle screen.
Users can stop the unit either during the warm-up time or during a manual
sample run.
1.
In the First Sample Screen #1, press the “A” and “B” buttons
simultaneously to display the Reset Sample 2 screen.
2.
In the Reset Sample 2 screen, press the “B” button (“more options”) to
display the Stop Sampling screen.
3.
In the Stop Sampling screen, press the “A” button (“Stop Sampling”
option) to stop the manual sample run and display the Sampling
Complete - Screen #1. This action will stop both the first and second
manual sample run (if you have started a second manual sample run).
Stopping During
Warm Up
To stop a manual sample run
during the warm-up time:
Stopping During
Sampling
To stop a manual sample run
during a sample run: