background image

Basic Operation 

Downloading Data 

Thermo Fisher Scientific 

PDM3700 Instruction Manual     

2-29

 

collect the shift data. If the downloaded data was part of a multi-unit 
shift, select yes on Multi-unit Shift field and enter the serial number of 
the first unit of the shift and the second unit of the shift. Ensure that 
the serial numbers of the two units are entered correctly and in the 
correct order. This information will be used to properly combine the 
data from the two samplers for the entire completed shift. The data 
from both samplers must be downloaded and sent to MSHA. 

9.

 

After entering the required information, you can view various graphs of 
the logging parameters that were recorded during the sample event. 
Select the parameters to view using the drop-down menu. For more 
information on the download data screen, refer to the “Download 
Graph Screen” section in Chapter 4. 

10.

 

After you have viewed the data, select the “Print Dust Data Card” 
button The Dust Data Card screen will display. Select the “Print” 
button to print a paper record of the sample run. This hard-copy 
record will display the data entered when the sample run was first 
programmed and the data results that were recorded. 

11.

 

After reviewing the data and to collect the data to send to MSHA for 
validation, select the Save Data button on the graph screen. When 
saving the data, you will be prompted to save the data to the PC. 
WinPDM will prompt to save two data files, one for local use and 
viewing, and a second data file for use by MSHA. The first data file is a 
.csv format that can be easily imported into Microsoft Excel, or other 
similar spreadsheet program. The second data file is a .MSHA file 
format and is required to be submitted to MSHA for validation. The 
.MSHA data file contains the data in the .csv file as well as a few 
additional information details that will be used by MSHA for 
validation. If the .MSHA data file is not required for a shift, press 
cancel when prompted to save the .MSHA data file. 

Note  

To ensure proper data storage and file information, refer to MSHA 

guidelines for data reporting requirements. 

  

12.

 

Select the “Save” button to save the data file. 

Note  

The WinPDM software automatically names the data files with the 

date and start hour of the sample run (yyyy-mm-dd-hh), the PDM3700 
unit serial number, the Mine ID number, and the vendor/instrument code. 
For example: 2014-08-19-06-0212502-0503836-A.csv or .MSHA. 

  

Summary of Contents for PDM3700

Page 1: ...Model PDM3700 Personal Dust Monitor Part Number 42 009904 3700 25Mar2016...

Page 2: ...lable in all countries Please consult your local sales representative for details TEOM is a registered trademark of Thermo Fisher Scientific Other trademarks are the property of their respective holde...

Page 3: ...h all costs prepaid by Buyer Replacement parts may be new or refurbished at the election of Seller All replaced parts shall become the property of Seller Shipment to Buyer of repaired or replacement P...

Page 4: ...Y OTHER THAN SELLER WITHOUT SELLER S PRIOR WRITTEN APPROVAL OR ANY USE OF REPLACEMENT PARTS NOT SUPPLIED BY SELLER SHALL IMMEDIATELY VOID AND CANCEL ALL WARRANTIES WITH RESPECT TO THE AFFECTED PRODUCT...

Page 5: ...ientific has contracted with one or more recycling disposal companies in each EU Member State and this product should be disposed of or recycled through them Further information on Thermo Fisher Scien...

Page 6: ...umes all liability associated with the use of this instrumentation The seller further disclaims any responsibility for consequential damages Use of this product in any manner not intended by the manuf...

Page 7: ...packs shall take place while installed in the PDM3700 and in fresh air in Battery Charging stations as defined in 30 CFR 75 340 or fresh air above ground Instrument This product is certified as intri...

Page 8: ...61000 3 3 95 A1 02 4 dmax Electrical Fast Transients EN 61000 4 4 2004 1kV AC mains 1kV I O Surges EN 61000 4 5 95 A1 01 0 5kV Line to earth 1kV Line to protected earth Line Conducted RF Susceptibili...

Page 9: ...ientific PDM3700 Instruction Manual v Equipment Rating The following information can be used to determine the power service requirements of this product Line Voltage 120 V 60 Hz 20 Amp 240 V 50 Hz 10...

Page 10: ...s ignored Safety and Equipment Damage Alerts in this Manual Alert Description DANGER Misusing or mishandling the battery pack can lead to fluid leakage heat generation fire or an explosion WARNING The...

Page 11: ...N PDM3700 maintenance should only be performed by trained authorized personnel CAUTION Refer to Section 1 for a comprehensive list of cautions and warnings for the battery pack assembly CAUTION Always...

Page 12: ...nual Thermo Fisher Scientific NIOSH Certification Label This label is affixed to the back of the instrument Only the parts listed should be used in order to maintain certifications Note All labels mus...

Page 13: ...12 Idle Screen 2 14 Starting a Primary Sample Run 2 14 Start Sampling Screen 2 15 Exit Screen 2 16 Start Sample Screen 2 16 Starting a Manual Sample 2 16 Starting a Second Sample Run 2 17 Second Sampl...

Page 14: ...rs 3 14 Program Shift Setup Problems 3 16 Correcting Shift Setup Problems 3 17 Viewing and Saving Data 4 1 Configuring Data Downloading 4 2 Downloading Data 4 3 Download Graph Screen 4 8 Dust Data Car...

Page 15: ...ion Manual xi Performing a Flow Calibration 5 33 Troubleshooting 6 1 Start Up Diagnostics 6 1 Instrument Diagnostics 6 2 Leak Testing 6 4 Sample Path Instrument Plumbing 6 5 Case Battery Compartment 6...

Page 16: ......

Page 17: ...s or other occupational environments with elevated particulate concentrations The PDM3700 is a wearable device that sets up quickly delivers accurate results recharges quickly and is easily serviced T...

Page 18: ...condensing Altitude 3 500 to 10 000 feet Temperature Sampled Air 5 to 105 F 20 to 40 C typical range is 32 85 F 0 30 C Temperature Shock 40 C min Temperature Charging 32 105 F 0 to 40 C The PDM 3700...

Page 19: ...the same charger that was supplied with the PDM3600 The PDM3600 included the integrated cap lamp in the device and needed a charging circuit to charge that battery This charger circuit was removed fro...

Page 20: ...TEOM filter the batteries allow a 12 hour run up to a loading of 5 mg User definable manual and programmed sample collection options based on worker shift lengths Capable of indoor and outdoor use On...

Page 21: ...oviding information about internal flows sample inlets and the flow control calculations Chapter 2 Basic Operation describes how to set up the hardware and software for the PDM3700 and provides an ove...

Page 22: ...tapered element through the air temperature and relative humidity sensor block past the flow differential pressure sensor and finally through the pump The air stream then exits the system The PDM3700...

Page 23: ...h air environment The lithium ion battery packs are designed to meet all U S Department of Transportation DOT criteria per UN Manual of Tests and Criteria Part III sub section 38 3 T1 T2 T3 T4 T5 T7 V...

Page 24: ...not allow them to get wet Do not break open or damage the battery packs or the casing of the lithium ion batteries Do not pierce the battery packs with a nail or other sharp object Do not strike it wi...

Page 25: ...ied individuals may perform maintenance calibration and sampling pursuant to 30 CFR Parts 70 71 90 This includes changing the battery packs sample pump cleaning the sample inlet and other maintenance...

Page 26: ......

Page 27: ...ple runs Upon receiving the PDM3700 unit ensure that a TEOM filter is installed in the mass transducer and connect the unit to the charger Figure 2 1 To fully charge the internal lithium ion batteries...

Page 28: ...ed CD ROM WinPDM software While the PDM3700 may be used as a stand alone device many of its features require the use of additional equipment To access the full functionality of the PDM3700 the followi...

Page 29: ...the belt loops remove the sealing screws installed on the charger side of the instrument and using the provided belt loop screws attach the belt loops to the side of the instrument When attaching the...

Page 30: ...nsate when the vertical orientation of the sampler changes If the sampler is placed or worn in a position where the display is upside down the sampler will detect this position and discard the real ti...

Page 31: ...first sample run Charging must be completed in a fresh air environment 1 Locate the PDM3700 charger and its power cord attach the power cord to the back of the charger and then install the charger s...

Page 32: ...nector Figure 2 6 and Figure 2 7 Figure 2 6 Setting the charger connector onto the unit Figure 2 7 Locking in the PDM3700 charger connection 4 Install the 15 pin connector of the communications power...

Page 33: ...DM3600 included the integrated cap lamp in the device and needed a charging circuit to charge that battery This charger circuit was removed from the charger supplied with the PDM3700 The charger suppl...

Page 34: ...d only by MSHA certified individuals pursuant to 30 CFR Parts 70 71 90 All charging of these battery packs shall take place while installed in the PDM3700 and in fresh air in Battery Charging stations...

Page 35: ...fork of the filter exchange tool under the TEOM filter that is in the mass transducer so that the filter disk is between the lower fork and the upper tab of the filter exchange tool The tines of the...

Page 36: ...lter installed on the tapered element TE but the filter must be changed daily or with each shift 1 Clean the exchange tool with canned air or a clean cloth Use it to pick up a new TEOM filter from the...

Page 37: ...the spring loaded filter exchange tool on top of the filter and apply downward pressure until the filter tool stops Figure 2 17 and Figure 2 18 4 Rotate the mass transducer 90 and press again Repeat...

Page 38: ...30 seconds of inactivity the PDM3700 will then automatically turn off its LED display If the PDM3700 is connected to the WinPDM software the PDM3700 display will stay on continuously The display on t...

Page 39: ...Basic Operation PDM Display Thermo Fisher Scientific PDM3700 Instruction Manual 2 13 Figure 2 20 First manual sample run screen hierarchy...

Page 40: ...instrument s serial number XXXXXXX Ver X XX This field contains the version number X XX of the system firmware A B action menu Press the A and the B button at the same time to display the unit s Start...

Page 41: ...nstrument s display and buttons When in the Idle screen or the Sampling Complete Screen 1 if a sample has been recently completed press both the A and the B buttons simultaneously to display the Start...

Page 42: ...an initiate a manual sample run through the Start Sample screen When in the Start Sampling screen Figure 2 23 press the B button to display the Start Sample screen Figure 2 25 Figure 2 25 Start Sample...

Page 43: ...button to begin a manual sample run The Warming screen will display Figure 2 26 When the sample run begins the instrument will display First Sample Screen 1 Figure 2 27 Figure 2 26 Warming screen Figu...

Page 44: ...d contains the averaged mass concentration value mg measured during the previous 15 minutes CUM2 CONC This field contains the cumulative mass concentration value mg m3 measured from the beginning of t...

Page 45: ...user can stop a manual sample run using the Stop Sampling screen Figure 2 31 Stopping a sample run will stop the main sample run as well as the second sample run if one has been started Stop the samp...

Page 46: ...the A and B buttons simultaneously to display the Stop Sampling screen 2 In the Stop Sampling screen press the A button Stop Sampling option to stop the manual sample run and display the Idle screen...

Page 47: ...ation Chapter 4 Warming 28 07 This field contains the time countdown minutes seconds of the PDM3700 unit s warm up time duration before the unit starts a first primary sample run At the end of the war...

Page 48: ...he shift for the first primary sample run Percent of exposure limit is the percentage of exposure limit for the first primary sample run Figure 2 34 First Sample Screen 2 The First Sample Screen 3 Fig...

Page 49: ...un the PDM3700 will run automatically until the sample cycle ends After 30 seconds of inactivity at the end of the sample run the PDM3700 will turn off its LED display Press the A button to reactivate...

Page 50: ...wing section Status Conditions for more information on status codes After 30 seconds of inactivity in the Sampling Complete Screen 1 the PDM3700 will turn off its LED display Press the A button to rea...

Page 51: ...ample Screen 1 press the A and B buttons simultaneously to display the Reset Sample 2 screen 2 In the Reset Sample 2 screen press the B more options button to display the Stop Sampling screen 3 In the...

Page 52: ...during a programmed or primary sample run A blinking S in the upper right corner of the display indicates that a status condition has been encountered The blinking can be stopped by navigating to the...

Page 53: ...ion is reported when the PDM3700 unit detects an instantaneous significant and sustained change in the total mass value It is designed to indicate that an event has occurred that may invalidate the cu...

Page 54: ...nditions during the sample run the Status Code List screen will display 6 If the status screen displays select the Save As button The Save As screen will display Accept the data file name that the sof...

Page 55: ...a entered when the sample run was first programmed and the data results that were recorded 11 After reviewing the data and to collect the data to send to MSHA for validation select the Save Data butto...

Page 56: ...3 If you do not want to save the file select the Cancel button The Download Graph screen will display 14 Click Cancel if you wish to exit the Download Graph screen 15 After saving the data please refe...

Page 57: ...delivered however this may not be the most current version of the software available Always check that you are using the latest MSHA approved version prior to programming the PDM3700 and downloading t...

Page 58: ...installation program Then the WinPDM InstallShield Wizard screen with a Welcome message will display Select the Next button Figure 3 1 InstallShield Wizard screen with Welcome message 5 The InstallShi...

Page 59: ...the WinPDM program Refer to Chapter 2 for information on connecting the PDM3700 unit to the charger and establishing communications 1 Connect the PDM3700 unit to its charger and the charger to your PC...

Page 60: ...PDM3700 and the PC is successful the unit s serial number will display in the top right hand corner of the screen Figure 3 3 Connect screen 6 If the connection is not successful or if there is no inst...

Page 61: ...s complete users can disconnect the PDM3700 and exit the WinPDM software 1 When in the PDM COM1 screen select the Done button located in the bottom right hand corner of the screen The WinPDM Main scre...

Page 62: ...C and ensure that the charger and the PDM3700 are both functioning correctly 2 Start the WinPDM software application 3 When in the PDM COM1 screen select the Program Shift button located on the left h...

Page 63: ...application and disconnect the PDM3700 unit from its charger If you leave the PDM3700 connected to the charger the PDM COM1 screen will display the Start Time date and time on the screen Figure 3 7 u...

Page 64: ...Using the WinPDM Software Starting a Programmed Sample Run 3 8 PDM3700 Instruction Manual Thermo Fisher Scientific Figure 3 7 Warming Figure 3 8 Running...

Page 65: ...he sample run using the WinPDM software 1 Connect the PDM3700 unit to its charger and the PC and ensure that the charger and the PDM3700 are both functioning correctly 2 Start the WinPDM software appl...

Page 66: ...ot be able to display a different firmware screen when the Programmed Start Time screen is visible on the PDM3700 unit Figure 3 10 Programmed start time PDM3700 screen The Programmed Start Time screen...

Page 67: ...screen will display Note After 30 seconds of inactivity in the Reset Sample 2 screen the PDM3700 will automatically display the First Sample Screen 1 When in the Program Shift screen Figure 3 12 you...

Page 68: ...t have an entry Figure 3 13 For example N A can be used if a field is not required Figure 3 13 Field Entry Error message The instrument stores the field information from the last programmed shift This...

Page 69: ...minute warm up time The warm up time automatically adjusts when the sample time is set 8 Sample Time This field contains the time duration hh mm of the sample run 9 Type of Sample select one This fiel...

Page 70: ...e Program Shift screen Figure 3 12 2 When in the Program Shift screen enter the basic parameters including if necessary the 8 digit identification number of the person who is wearing the PDM3700 unit...

Page 71: ...perator that will be using the PDM3700 unit if necessary in the box under 11 Occ Code 8 Use the Average Expected Temperature drop down menu Figure 3 16 to select the proper temperature setting for the...

Page 72: ...Start Time date and time on the screen until 35 minutes before the start date time At the 35 minutes mark the PDM COM1 screen will display a Warming message on the screen for 35 minutes After the war...

Page 73: ...the Program Shift screen identify the appropriate section of the screen that you must correct For example during warm up a sample could be redesignated from one wearer ID to another wearer ID on a dif...

Page 74: ......

Page 75: ...nd the RS232 cable to download data from the PDM3700 unit Two data files are available for download the first is a file with a comma separated value csv file format The second file is in msha format w...

Page 76: ...d date hh mm ss weekday mm dd yyyy ending time and date hh mm ss weekday mm dd yyyy and size of the data file Bytes Each time that you perform a sample run the PDM3700 unit logs the sample data and di...

Page 77: ...ger and the charger to the PC Ensure that the charger and the PDM3700 are both functioning correctly 2 Start the WinPDM software application and select the Download Sample button 3 When in the Downloa...

Page 78: ...0 Instruction Manual Thermo Fisher Scientific Figure 4 4 Download Data screen with progress bar 5 If the PDM3700 unit encountered any status conditions during the sample run the Status List screen wil...

Page 79: ...hift Information dialog box will appear Enter the tonnage for the downloaded shift along with the 30 day average for that shift Also enter the Certified Person MIIN WinPDM will not allow you to move f...

Page 80: ...For more information on the download data screen refer to the Download Graph Screen section later in this chapter 10 After you have viewed the data you can print the Dust Data Card Figure 4 9 on page...

Page 81: ...2 0503836 A csv or msha 13 A Save As dialogue box will display prompting the operator to save a csv file This file is for the use of the operator Select Save 14 Next a second Save As dialogue box will...

Page 82: ...elect the data parameter you want to view from a drop down menu When in the Downloaded Logging Parameters drop down menu select the downloaded logging parameter that you want to display on the graph F...

Page 83: ...d Graph screen select the Print Dust Data Card button located in the bottom center section of the screen to display the Dust Data Card screen When in the Dust Data Card screen you can print the Dust D...

Page 84: ...ch as Microsoft Excel open the data file Figure 4 10 At the top of the data file will be the data entered into the Program Shift screen when you first programmed the sample run Below the program shift...

Page 85: ...ing Data Viewing Data Files Thermo Fisher Scientific PDM3700 Instruction Manual 4 11 Figure 4 10 Excel file with downloaded data Program shift information Header footer information Data values of logg...

Page 86: ...folder The file format is a comma separated value csv text file format Using a spreadsheet program such as Microsoft Excel open the status list file Figure 4 11 The file will contain the time that the...

Page 87: ...The status file format is a comma separated value csv text file format Using a spreadsheet program such as Microsoft Excel open the data file Figure 4 12 You can locate the status code in the second c...

Page 88: ......

Page 89: ...e or as necessary Clean the mass transducer and sample tubing Replace the TEOM filter Section 2 and clean the mass transducer area and sample tube after each use or as necessary Monthly Perform a flow...

Page 90: ...uipment Damage Do not use acetone to clean any of the parts of the PDM3700 Several maintenance procedures including changing the filter cleaning the grit pot and cleaning the sample lines require the...

Page 91: ...3 When ready reinstall the mass transducer into the PDM3700 by aligning the flat side of the transducer toward the center of the PDM3700 and pushing it in until it seats firmly in place Figure 5 3 and...

Page 92: ...ned air Alcohol swab Silicone lubricant 1 Remove the mass transducer 2 Locate the grit pot on the right hand side of the PDM3700 unit Figure 5 5 Figure 5 5 Photo showing location of grit pot 3 Gently...

Page 93: ...he cyclone nozzle to clean the inside of the nozzle Use an alcohol swab to clean the inside and outside surfaces of the cyclone nozzle Figure 5 7 Figure 5 7 Instrument Inlet and Cyclone Nozzle 6 Reins...

Page 94: ...to meet agency requirements 35 37 The tubing must be checked prior to use each day and may be trimmed due to breakage or fraying but must always be between 35 37 during use If the tubing is shorter th...

Page 95: ...Use canned air then an alcohol swab to clean all surfaces of the mass transducer 4 Use canned air and an alcohol swab to clean the bell shaped inlet inside the PDM3700 unit and then direct a stream o...

Page 96: ...ean the bell shaped inlet with air 5 Use silicone lubricant to lightly coat o rings on bell shaped inlet and return tube 6 Direct a stream of air through the end of the sample line removed from the in...

Page 97: ...Calibration Cleaning the Mass Transducer and Sample Lines Thermo Fisher Scientific PDM3700 Instruction Manual 5 9 Figure 5 12 Direct a stream of air into the sample inlet Figure 5 13 Loosen screws on...

Page 98: ...ntenance and Calibration Cleaning the Mass Transducer and Sample Lines 5 10 PDM3700 Instruction Manual Thermo Fisher Scientific Figure 5 14 Open sample inlet assembly bracket Figure 5 15 Slide out inl...

Page 99: ...nance and Calibration Cleaning the Mass Transducer and Sample Lines Thermo Fisher Scientific PDM3700 Instruction Manual 5 11 Figure 5 16 Remove sealing screw using inlet spanner tool Figure 5 17 Clean...

Page 100: ...Install a new TEOM filter onto the mass transducer and install the mass transducer into the PDM3700 unit 8 Reassemble the sample inlet and reattach the sample line to the sample inlet 9 Install the sa...

Page 101: ...requires the supplied tubing adapters Other audit devices may need other adapters or no adapters at all Modify these instructions as necessary based on the audit device being used 1 Disconnect the sam...

Page 102: ...is within the acceptable criteria of 0 055 lpm end the sample audit run remove the tubing adapter and reinstall the sample line If the flow meter is not within your acceptance criteria a flow calibra...

Page 103: ...a sample line leak check Once a year the calibration constant K0 number of the unit should be audited Tool materials needed for this procedure K0 audit kit 59 010020 Alcohol swab Small tweezers Note...

Page 104: ...l tweezers to add the weight to the filter Figure 5 22 Prompt to add weight 5 The computer will prompt to add three more weights Add each weight consecutively as shown in Figure 5 23 Figure 5 23 Add f...

Page 105: ...r is placed or worn in a position where the display is upside down the sampler will detect this position and discard the real time mass and mass concentration data until this situation is corrected Wh...

Page 106: ...e the TE frequency 4 When stable the computer will prompt to position the instrument upright as shown in Figure 5 26 Place on a vibration free table and support instrument so it does not move during t...

Page 107: ...Sensor Thermo Fisher Scientific PDM3700 Instruction Manual 5 19 Figure 5 27 Position the instrument on the side as shown 6 When complete the computer will indicate a pass or fail status Figure 5 28 F...

Page 108: ...0 All charging of these battery packs shall take place while installed in the PDM3700 and in fresh air in Battery Charging stations as defined in 30 CFR 75 340 or fresh air above ground WARNING Do not...

Page 109: ...s the case leak check prior to operating the instrument Correct the leak and re test the instrument if the leak check does not initially pass Do not operate the instrument if the system does not pass...

Page 110: ...nual Thermo Fisher Scientific Figure 5 29 Cover panel screws Figure 5 30 Screws on top display Equipment Damage Do not attempt to remove the security non Phillips head screws on the top or front of th...

Page 111: ...he leak and re test the instrument if the leak check does not initially pass Do not operate the instrument if the system does not pass the case leak check 1 Inspect the gasket and sealing surface to e...

Page 112: ...er step Press down on back of cover panel aligning holes 3 and 4 Put a screw into either location 3 or 4 and again tighten until snug Torque the screws to 6 in lbs respectively on location 1 and then...

Page 113: ...PDM3700 case is opened for maintenance you must perform a case leak check after reassembling the case The system must pass the case leak check prior to operating the instrument Correct the leak and r...

Page 114: ...lug in open end of sample tubing Figure 5 36 Figure 5 36 Insert sample line plug into tubing 5 Prior to the start of the leak check locate the pump exhaust muffler Figure 5 37 When the sample path lea...

Page 115: ...aintenance and Calibration Performing Sample Path and Case Leak Checks Thermo Fisher Scientific PDM3700 Instruction Manual 5 27 Figure 5 37 Pump exhaust muffler Figure 5 38 Placing the leak check putt...

Page 116: ...ce the inlet adapter in the inlet message Install the case leak check suction cup into the sample line Figure 5 39 Select OK Figure 5 39 Leak check suction cup installed in sample line 8 Another Warni...

Page 117: ...700 unit will perform the case leak check and display a pass or fail message on the Calibrate Audit screen Figure 5 42 Remove the leak check putty from the exhaust muffler Figure 5 42 Leak check scree...

Page 118: ...air above ground WARNING Whenever the PDM3700 case is opened for maintenance you must perform a case leak check after reassembling the case The system must pass the case leak check prior to operating...

Page 119: ...0 unit Press the small tab on the wire connection to disconnect the battery wires Figure 5 44 and Figure 5 45 Equipment Damage When moving the battery pack to access the wire connections be sure to mi...

Page 120: ...ferred method of disposal Contact Thermo Fisher Scientific for further information Do not dispose of the battery packs in fire or heat CAUTION Always properly dispose of the PDM3700 battery packs recy...

Page 121: ...and Figure 5 47 Figure 5 46 Mass transducer chamber Figure 5 47 Mass transducer O rings 3 Using the cotton swab apply the silicone compound to the O rings 4 Install the PDM3700 cover panel refer to th...

Page 122: ...ample line from the sample inlet assembly 3 Perform a sample path leak check 4 Connect the PDM3700 unit to its charger and start the WinPDM software 5 When in the PDM COM1 screen select the Flow Calib...

Page 123: ...gure 5 49 Install tubing adapter into the sample line 9 Select the OK button The Flow Calibration screen will display with the Pump Speed box active Figure 5 50 Figure 5 50 Flow calibration screen wit...

Page 124: ...s to ensure the flow meter is stable 12 Select the Apply button The PDM3700 unit will begin the flow calibration routine and will display a Please wait gathering data message on the Flow Calibration s...

Page 125: ...adjust the pump speed Enter the Orifice 1 flow rate from the flow meter typically 2 0 lpm 18 Select the Apply button The PDM3700 unit will continue the flow calibration routine and will display a Ple...

Page 126: ...d white box active 23 Do Not adjust the pump speed Enter the Orifice 2 flow rate from the flow meter typically 1 8 lpm 24 Select the Apply button The PDM3700 unit will continue the flow calibration ro...

Page 127: ...utton to keep the old values Press the Restart button to perform a new calibration or the Cancel button to exit 27 Remove the red orifice from the sample tubing and flow meter tubing 28 Install the sa...

Page 128: ......

Page 129: ...ILURE To view the cause of the fault press the A and B buttons press B to scroll until the View Failures options is displayed Press A to view the failure s The flashing DIAGNOSTIC FAILURE alert will s...

Page 130: ...m starting Below is a list of the possible diagnostic failures along with the associated WinPDM Instrument Diagnostic PDM3700 Diagnostic Failure WinPDM Instrument Diagnostics AIR TEMP FAILURE AIR TEMP...

Page 131: ...R TEMP Defective Disconnected Sensor Check connection AMBIENT PRES Defective Sensor Contact TFS service AMBIENT TEMP Defective Disconnected Sensor Check connection CLOCK Defective Component Contact TF...

Page 132: ...ed Install Mass Transducer Check for filter Contact TFS service NO MASS TRANSDUCER Mass Transducer not installed Install Mass Transducer RH SENSOR Defective Sensor Contact TFS service RH TEMP Defectiv...

Page 133: ...nstruction Manual 6 5 The most common sources of a leaking sample path in order of most likely to cause leak Grit Pot not seated needs lubrication torn or punctured Mass transducer o rings needs silic...

Page 134: ...Thermo Fisher Scientific Cyclone gasket screws not tight gasket damaged metal plates damaged If there is a problem with the attachment of the cyclone plate and gasket assembly contact Thermo Fisher Sc...

Page 135: ...of a leaking battery compartment in order of most likely to cause a leak Case screws screws missing not tightened to 6 in lbs per the sequence shown in the following illustration Cover Panel cracked o...

Page 136: ......

Page 137: ...00 Charger Parts Interface assembly charger PDM3700 55 010924 0001 Cable RS232 m f 10 ft 07 000587 Fuse 2A time delay 04 003268 Cord power detachable for charger input power 07 000593 Charger assembl...

Page 138: ...721 0003 Inlet bracket clip attachment 55 012278 Sample tubing 32 006785 0001 Sample tubing 50 ft roll 32 006785 0050 Tubing clip 30 012091 Inlet spanner tool 30 012264 Recommended spare parts to have...

Page 139: ...00 Instruction Manual A 3 Figure A 1 Sample Inlet Assembly Figure A 2 Completed Sample Tube and Sample Inlet Assembly Inlet Bracket 38 012094 Inlet Bracket Clip Attachment 55 012278 Sealing Screw 35 0...

Page 140: ......

Reviews: