EN-75
Thermo Scientific Orion Star
™
and Star Plus Meter User Guide
Setup Information for Interfacing Meter, Autosampler Unit
and Windows
®
Hyperterminal
1 . Select a location for the system and unpack the autosampler and meter
components . Prepare the autosampler according to the autosampler user guide .
2 . Interface the autosampler with the meter . Use the autosampler to meter
interface cable, Cat . No . ATA02 (labeled 254210-001), to connect the 2 .5
mm phono jack RS232 port on the meter to the Com 1 RS232 port on the
autosampler .
3 . Use special NULL modem cable (provided with the autosampler) and connect
COM 2 of autosampler to COM port of computer .
Note:
Make sure your Hyperterminal COM settings are set to the same settings
as COM2 of the autosampler unit.
4 .
Adjust DIP switch 4 and DIP switch 5 to the OFF position (up) to set the
autosampler to 9600 baud for Com 1 .
5 . Prepare and install the electrodes . Prepare the electrodes according to the
electrode user guides . Connect the electrodes to the back of the meter
using the extension cables and cable management system provided with the
autosampler . Insert the electrodes into the electrode holder on the autosampler .
If a stirrer probe will be used, connect the stirrer probe to
the stirrer jack on the
back of the meter and insert it into the electrode holder on the autosampler .
6 . Connect the meter and autosampler to a power supply and turn on the power .
7 . Verify the baud rate for the meter is 9600 . Adjust the meter setting if necessary .
8 . Enter the AutoSampler setup menu in the meter and set to ON . Follow
the instructions in the section titled “Meter Preparation for Operating the
Autosampler” .
Orion Star Plus Benchtop Meter and Autosampler Interfacing