Thermo Scientific Orion Star
™
and Star Plus Meter User Guide
EN-16
AUTO-READ
™
, Continuous or Timed Measurement Settings
• In the AUTO-READ mode, the meter starts taking
a measurement when
is pressed . Once the
measurement is stable, the display freezes and the data
is logged and printed . The AUTO-READ mode also controls
the stirrer . The stirrer starts when
is pressed and stops
when the measurement becomes stable .
• In the continuous mode, the meter is constantly taking measurements and
updating the display . Press
to log and print a measurement in this mode .
• In the timed mode, the meter is constantly taking measurements and updating
the display . The meter logs and prints the measurement at the selected time
interval . Timed dissolved oxygen measurements with the RDO probe are taken
only at the selected time interval, which conserves the meter battery power .
1 . In the measurement mode, press
.
2 . Press
/
to scroll through the setup menu until
is displayed
on the top line .
3 . Press
to accept the selection and move the arrow icon to the middle line .
4 . If the timed measurement setting is active, press
/
to scroll
through
for the measurement read type and
for the timed
reading interval .
5 . Press
to accept the selection and move the arrow icon to the bottom line .
6 . To scroll through a list of options on the bottom line, press
/
until the
desired option is displayed . To enter a numeric value for an option on the bottom
line, press
/
to adjust each digit and
to move to the next digit .
7 . Press
to accept the selection and move the arrow icon to the top line .
8 . Repeat steps 3 through 7 to change another measurement setting or press
to return to the measurement mode .