Thermo Scientific Orion Star
™
and Star Plus Meter User Guide
EN-70
Cat. No.
011005
011006
011007
01100910
011008
Temperature
(°C)
111.9 mS/cm
Conductivity
Standard
(mS/cm)
12.9 mS/cm
Conductivity
Standard
(mS/cm)
1413 µS/cm
Conductivity
Standard
(µS/cm)
147 µS/cm
Conductivity
Standard
(µS/cm)
100 µS/cm
Conductivity
Standard
(µS/cm)
0
65.10
7.135
776
81
54
1
66.84
7.344
799
83
56
2
68.59
7.555
822
86
58
3
70.35
7.768
846
88
59
4
72.12
7.983
870
91
61
5
73.91
8.200
894
93
63
6
75.70
8.418
918
96
64
7
77.50
8.638
943
98
66
8
79.32
8.860
968
101
68
9
81.15
9.084
992
103
70
10
82.98
9.309
1017
106
72
11
84.83
9.535
1043
108
73
12
86.69
9.763
1068
111
75
13
88.56
9.993
1094
114
77
14
90.45
10.22
1119
116
79
15
92.34
10.46
1145
119
81
16
94.24
10.69
1171
122
83
17
96.15
10.93
1198
125
85
18
98.08
11.16
1224
127
87
19
100.0
11.40
1251
130
88
20
102.0
11.64
1277
133
90
21
103.9
11.88
1304
136
92
22
105.9
12.12
1331
138
94
23
107.9
12.36
1358
141
96
24
109.9
12.61
1386
144
98
25
111.9
12.85
1413
147
100
26
113.9
13.10
1441
150
102
27
115.9
13.35
1468
153
104
28
117.9
13.59
1496
156
106
29
120.0
13.84
1524
159
108