EN-31
Thermo Scientific Orion Star
™
and Star Plus Meter User Guide
Chapter IX
RDO
®
Optical Dissolved
Oxygen Technique
Note:
The RDO optical dissolved oxygen meters have serial numbers that
begin with R (i.e. R12345). Only the RDO meters are compatible with the
RDO optical probes.
Optical Cap Overview
The RDO optical probe has an internal clock that counts down the 365 day
lifespan of a new optical cap . The countdown begins when the optical cap is
installed on the RDO probe, the probe is connected to the meter and the first
measurement is taken . The 365 day countdown cannot be reset or changed once
the first measurement is taken . Each optical cap has a unique serial number that is
recognized by the RDO probe, so reinstalling the cap will not reset the countdown .
Note:
The power to the meter must be turned off when a new optical cap
is installed on the RDO probe. Once the optical cap is installed, turn the power
to the meter on and the new cap information will be sent to the meter.
To print the optical cap information:
1 . Connect the meter to a printer or computer and verify the meter baud rate
and output settings in the setup menu .
2 . From the measurement mode, press
.
3 . Press
/
to scroll through the setup menu until
is displayed
on the top line and
is displayed on the middle line .
4 . Press
two times to move the arrow icon to the bottom line .
5 . Press
/
until
is displayed on the bottom line .
6 . Press
to print the optical cap information . Press
to return to the
measurement mode .