Thermo Scientific Orion Star
™
and Star Plus Meter User Guide
EN-80
Pausing the Autosampler
To pause the autosampler during operation, press
. To resume operation, press
again .
Stopping the Autosampler
To stop the autosampler during operation and abort the run, press and hold
for about 3 seconds .
Moving the Electrode Arm
At the end of a run, when the electrode arm
has returned to home position and is lowered
into beaker one, press
to lift the electrode
arm and press
again to lower the electrode
arm . See the diagram to the right . This action
is operational only when the autosampler is
not running and the electrode arm is in the
home position .
Turning Off the Autosampler Function
To disable the autosampler, press
in the measurement mode to enter the
setup menu . Press
until
is displayed on top line . Press
until the
arrow icon points to the middle line and press
until
is displayed . Press
until the arrow icon points to the bottom line and press
until
is
displayed . Press
until the arrow icon points to the top line . Press
to return
to measurement mode .
Autosampler Dip Switch Settings
The dip switch cover is located on
the rear panel of the autosampler and
a small Phillips head screw driver is
needed to remove the dip switch cover .
The ON position for the dip switches is
down and the OFF position is up . To set
the autosampler dip switches to 9600
for Com 1, adjust DIP switch 4 and DIP
switch 5 to the OFF position (up) .
DIP 1
DIP 2
DIP 3
DIP 4
DIP 5
Even Parity
ON
No Parity
OFF
9600 Baud
OFF
OFF
1200 Baud
ON
OFF
2400 Baud
OFF
ON
38500 Baud
ON
ON
Orion Star Plus Benchtop Meter and Autosampler Interfacing