EN-41
Thermo Scientific Orion Star
™
and Star Plus Meter User Guide
a . Automatic calibration (for conductivity standards of 10 μS/cm, 1413 μS/cm
and 12 .9 mS/cm) – When the
µS/cm
or
mS/cm
icon stops flashing, the
meter will display the temperature-corrected conductivity of the standard .
b . Direct calibration – When the
µS/cm
or
mS/cm
icon stops flashing,
enter the conductivity value of the standard at the conductivity reference
temperature . (Refer to page EN-12 .) Press
until the first digit to be
changed is flashing, press
/
to change the value of the flashing digit
and continue to change the digits until the correct value is displayed . Once
the value is set, press
until the decimal point is in the correct location .
5 . Press
to proceed to the next calibration point, rinse the conductivity probe
with distilled or deionized water, place it into the next conductivity standard
and repeat step 4a / 4b or press
to save and end the calibration .
6 . The cell constant will be displayed in the main field and the meter will
automatically advance to the measurement mode .
Conductivity Technique
Summary of Contents for Orion Star
Page 367: ......