Maintaining & Troubleshooting
Thermo Fisher Scientific
Maintaining & Troubleshooting
4-25
REC-4105 Rev L
If the displays do not light, the TMD has been damaged and must
be replaced.
If the fuse tested bad in the first procedure:
1. Replace the fuse with the correct sized fuse.
2. Turn the power switch back on.
3. If the displays do not light, turn the power switch off.
4. Test the fuse again.
If the fuse tests good, follow Steps 8–10 in the first test procedure.
If the fuse tests bad, continue with this procedure.
5. Disconnect all wiring to the CPU board except the power on TB1.
6. Replace the fuse with the correct-sized fuse.
7. Turn the power switch back on.
8. If the displays do not light, turn the power switch off.
9. Test the fuse again.
If the fuse tests good, follow Steps 8–10 in the first test procedure.
If the fuse tests bad, the TMD has been damaged and must be replaced.
The previous procedure should leave you with a functioning display
with no field wiring except incoming power. The next test is to
determine whether a peripheral system component may be causing the
fuse to blow.
You will be connecting components to the CPU board one at a time until
the fuse blows. Refer to the field wiring diagram for your system or the
Field Wiring Diagram (D07344A-E001) in Appendix D for correct
wiring connections.
Summary of Contents for Oretronic III
Page 4: ......
Page 8: ......
Page 14: ......
Page 68: ......
Page 100: ......
Page 114: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 134: ......
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ......
Page 139: ......
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 147: ......
Page 148: ......
Page 149: ......
Page 150: ......
Page 151: ......
Page 152: ......