A-8
Thermo Fisher Scientific
REC-4105 Rev L
R E G I S T E R
A D D R E S S
A C C E S S T Y P E
D E S C R I P T I O N
40032
Read Only
P
Indicates how many LEDs are lit on the bar graph.
Uni-polar mode:
M S B = 0 x 0 0
L S B = N u m b e r o f L E D b a r s l i t
In this mode, zero is at the bottom of the graph. LSB is proportional
to the absolute value of the metal signal.
Example: LSB = 3 indicates that three bars above the bottom are lit.
This can mean +3 or –3.
Bi-polar mode:
M S B = 0 x 0 1
L S B = N u m b e r o f L E D b a r s l i t
In this mode, zero is in the middle of the graph. LSB is proportional
to the metal signal. Sign is taken is taken into account.
Example: LSB = 3 indicates that seven bars below the center are lit.
This means –7.
40033
Read/Write
I
Audit fine sensitivity if Audit Check is active.
40034
Read/Write
I
Audit coarse sensitivity if Audit Check is active.
40035
Read/Write
I
Audit interval = Time, in hours, between Audit Checks.
40036
Read/Write
I
Filter = The noise filter used on the TMD.
0 - F i l t e r e m p h a s i z e s m i d d l e o f r e s p o n s e c u r v e
1 - F i l t e r e m p h a s i z e s b e g i n n i n g o f r e s p o n s e
c u r v e
2 - F i l t e r we i g h t s a l l o f t h e c u r v e e q u a l l y
40037
Read/Write
I
Frequency = The operating frequency of the TMD.
0 - 9 0 0 H z
1 - 7 8 0 H z
2 - 6 6 0 H z
3 - 1 0 2 0 H z
40038–40041
Read Only
P
Software version, stored as two characters per integer.
Left most character is stored in the MSB of Register 40038.
Summary of Contents for Oretronic III
Page 4: ......
Page 8: ......
Page 14: ......
Page 68: ......
Page 100: ......
Page 114: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 134: ......
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ......
Page 139: ......
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 147: ......
Page 148: ......
Page 149: ......
Page 150: ......
Page 151: ......
Page 152: ......