Installing the TMD
2-8
Installing the TMD
Thermo Fisher Scientific
REC-4105 Rev L
Follow this procedure to mount the control unit enclosure.
1. Open the enclosure door.
2. Remove the power cover to allow access to the line power terminals.
3. Bolt the enclosure to a flat, vertical surface using the 2-position
mounting feet on the back of the enclosure (see Figure 2-1).
4. Punch the required conduit holes in the bottom of the enclosure for
the power supply cable, coil cables, control wiring, communications
cable(s), and any additional signal wires and conduit.
5. Locate the holes in the appropriate areas of the enclosure to separate
high (greater than 30 V) and low (less than 30 V) voltages (see
Figure 2-2).
6. Install
conduit.
Figure 2–2.
Control Unit Enclosure - Bottom View
WARNING
Make sure the mounting surface is flat so that the
fibreglass enclosure does not twist or warp when the
mounting bolts are tightened.