C-
3
REC 4105 Rev L
2. Press
ENTER
3. Scroll through the SETUP menu until EST INT displays
4. Press
ENTER
The current interval displays.
5. Use the arrow keys to set the designated interval time between 1 and
24 hours.
Note
: When you set the calibration reading and the interval you must
scroll back through the menus until you reach the RUN menu. If you do
not scroll back to the RUN menu, your changes are not saved.
Last EST (EST LAST) displays the last signal reading for the last
interval EST test. Use the following procedures to display the last EST
value:
1. At the TMD control panel, from the RUN menu, scroll to the
CALIBRATION menu.
2. Press
ENTER
3. Scroll through the CALIBRATION menu until EST LAST displays.
4. EST reading from the last interval time displays.
EST Test (EST TEST) runs EST through a system test on demand. Use
the following procedures to run an EST test:
1. At the TMD control panel, from the RUN menu, scroll to the
CALIBRATION menu press the up arrow until the display toggles
to Y.
2. Press
ENTER
3. Scroll through the CALIBRATION menu until EST TEST displays.
4. Press
ENTER
Last EST (EST
Last)
EST Test (EST
Test)
Summary of Contents for Oretronic III
Page 4: ......
Page 8: ......
Page 14: ......
Page 68: ......
Page 100: ......
Page 114: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 134: ......
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ......
Page 139: ......
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 147: ......
Page 148: ......
Page 149: ......
Page 150: ......
Page 151: ......
Page 152: ......