Maintaining & Troubleshooting
4-26
Maintaining & Troubleshooting
Thermo Fisher Scientific
REC-4105 Rev L
The basic procedure for testing components is:
1. Turn the power switch (SW6) off.
2. Wire in a component.
3. Turn the power switch on.
4. Check the front panel.
If the displays are lit, that component did not blow the fuse. Repeat
the procedure for the next component.
If the displays do not light, the fuse is blown. The wiring for that
component is causing a problem.
Wire in the components, one at a time, in the following order:
• Transmitter
coil
• Clip detector A
• Clip detector B
• High pile detector
• SYNC
input
• Receiver
coil
• Remote
panel
• Alarm
outputs
If all of the system wiring tests all right, go on to the next procedure for
checking the TMD configuration.
1. Turn on power to the control unit (SW6).
2. Place the TMD in Setup/Calibrate mode.
3. Scroll to the MAT CODE display.
4. The value shown in the counter should be 3.
5. If the MAT CODE value is not 3, change it to 3 and press ENTER.