Maintaining & Troubleshooting
4-24
Maintaining & Troubleshooting
Thermo Fisher Scientific
REC-4105 Rev L
1. Turn on power to the control unit (SW6).
2. If the displays do not light, check power wiring on TB1.
3. If the line voltage setting is not correct for your detector (SW5),
correct it.
4. Turn the power switch (SW6) off.
5. Remove fuse (F1) and test it.
If the fuse tests good, put it back into the fuse holder and continue
with this procedure.
If the fuse tests bad, go to the next procedure.
6. Turn the power switch back on.
7. If the displays do not light, turn the power switch off and test the
fuse again.
8. Check the connectors on the CPU board and the ribbon cable to the
display board for correct mating on boards.
Re-insert any connector that may have come loose during shipping
or installation.
9. Check for coil or junction box damage using the procedures in
“Testing for Coil or Junction Box Damage”.
10. Turn the power switch back on.
the electronics cabinets, power supply cabinets,
control cabinets, or switch boxes.
WARNING
High voltage that may be present on leads could cause
electrical shock.
All switches must be OFF when checking input AC
electrical connections, removing or inserting printed
circuit boards, or attaching voltmeters to the system.
Use extreme caution when testing in, on, or around
the electronics cabinet, PC boards, or modules. There
are voltages in excess of 115 V or 230 V in these
areas.
Summary of Contents for Oretronic III
Page 4: ......
Page 8: ......
Page 14: ......
Page 68: ......
Page 100: ......
Page 114: ......
Page 131: ......
Page 132: ......
Page 133: ......
Page 134: ......
Page 135: ......
Page 136: ......
Page 137: ......
Page 138: ......
Page 139: ......
Page 140: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 147: ......
Page 148: ......
Page 149: ......
Page 150: ......
Page 151: ......
Page 152: ......